Jan 2, 2021 10:14
3 yrs ago
14 viewers *
English term

are added compatibility

English to Russian Tech/Engineering Electronics / Elect Eng
This test shall verified the compliance to IEC60601-1 ed. 3.1 electrical safety sections regarding leakage currents, Isolation barriers, defibrillation proof CF AP, Impedance of Protective Earth connections as difference tests to /IEC60601-1 test report/ (SAP DIS 300000319749) which was conducted by TÜV Süd for MA2 devices KS03468-02. Most of the isolation barrier distance measurements are still valid, only the barriers B and C have to be remeasured. Also residual voltage measurements on device KS03468-02 are still valid. The changes that can result in a difference of measured values are added compatibility of Single use scopes on the Office Line path and caging of the power supply to increase EMC stability.
Change log

Jan 2, 2021 10:14: changed "Kudoz queue" from "In queue" to "Public"

Proposed translations

17 mins
Selected

повышают соответствие

Внесенные изменения (которые могут привести к отличию измеренных значений) повышают совместимость указанных компонент, что улучшает их ЭМС-стабильность

--------------------------------------------------
Note added at 18 mins (2021-01-02 10:32:55 GMT)
--------------------------------------------------

Можно сказать и "добавляют совместимость" компонент
Something went wrong...
4 KudoZ points awarded for this answer.
8 mins

представляют собой добавочную совместимость

Something went wrong...