https://www.proz.com/kudoz.php/english-to-romanian/environment-ecology/3604142-so2-span.html

Glossary entry

English term or phrase:

SO2 span

Romanian translation:

gaz etalon/de calibrare SO2

Added to glossary by Claudia Coja
Dec 13, 2009 18:25
15 yrs ago
4 viewers *
English term

SO2 span

English to Romanian Tech/Engineering Environment & Ecology
"... 6. Switch the five-way ball valve MV1 on the required position (for a good calibration the first component must be SO2 span) to let the calibration gas flow into the analyzer. Switch the two-way ball valve MV2 on and regulate flow using nail valve on FL1 flowmeter at 150 l/h ..."

Multumesc anticipat pentru ajutor!!
Proposed translations (Romanian)
4 +3 (gaz de) span SO2
4 concentratia cunoscuta de SO2
Change log

Dec 13, 2009 20:22: cristina48 changed "Field (specific)" from "Mechanics / Mech Engineering" to "Environment & Ecology"

Dec 15, 2009 06:49: Claudia Coja changed "Edited KOG entry" from "<a href="/profile/641998">Ana-Maria Badea's</a> old entry - "SO2 span"" to ""gaz etalon/de calibrare SO2""

Discussion

Ana-Maria Badea (asker) Dec 14, 2009:
Draga Claudia, iti multumesc frumos, si eu am ajus la concluzie ca versiunea "gaz de calibrare" sau "gaz etalon" este buna ;) Multe multumiri inca o data tuturor!!
Claudia Coja Dec 14, 2009:
gaz de calibrare Ce zici de varianta "gaz de calibrare" ? http://www.spms.pub.ro/fisiere/depoluare/curs/cap04.pdf "sau alt gaz de calibrare adecvat"

cristina48 Dec 14, 2009:
interval/domeniu de măsurare ..."pt. o bună calibrare primul parametru introdus trebuie să fie intervalul/domeniul de măsurare al SO2". Cred totuşi că despre asta e vorba... De acord că apare în multe referinţe ca "gaz span" însă nu apare în dex şi nici nu am auzit în viaţa mea acest termen.
Claudia Coja Dec 14, 2009:
gaz etalon? "posibilitate de calibrare fara gaz etalon ..." http://caz.mecen.pub.ro/afriso.pdf
si in franceza suna similar " SO2 (Dioxyde de Soufre) - Bouteille gaz étalon à 5.0 ppm "http://www.accessoiresgaz.com/bouteille-gaz-etalon/bouteille...

Ana-Maria Badea (asker) Dec 14, 2009:
Din pacate nu este mentionat, am postat toata fraza si fragmentul referitor la SO2 span ...
cristina48 Dec 14, 2009:
important ar fi să ştim unitatea de măsură a acestui "SO2 span". În ce se măsoară?

Proposed translations

+3
36 mins
Selected

(gaz de) span SO2


vad ca si in RO e pastrat "span"
"Răspunsul la trecerea ***gazului span*** prin filtru.."
http://209.85.129.132/search?q=cache:zjUSuHPYNTAJ:www.anpm.r...
"gaz de span" http://www.mecrosystem.ro/produs/sistem-de-analiza-si-monito...



--------------------------------------------------
Note added at 1 day29 mins (2009-12-14 18:54:46 GMT)
--------------------------------------------------

In consens total cu Cristina, pentru ca si eu prefer sa traduc termenul atunci cand este posibil si sa-l preiau doar cand sunt in imposibilitate de a-i oferi un echivalent in limba romana, am mai cautat variante si am ajuns la concluzia ca si ***gaz etalon*** poate fi varianta corecta
"posibilitate de calibrare fara gaz etalon ..." http://caz.mecen.pub.ro/afriso.pdf
si in franceza suna similar " SO2 (Dioxyde de Soufre) - Bouteille gaz étalon à 5.0 ppm "http://www.accessoiresgaz.com/bouteille-gaz-etalon/bouteille...
Peer comment(s):

agree Sandra Roselee
1 hr
multumesc :)
agree Iosif JUHASZ
12 hrs
Multumesc :)
agree cristina48 : M-am mai documentat şi se pare că asta e!
20 hrs
multumesc, Cristina :)
Something went wrong...
4 KudoZ points awarded for this answer. Comment: "Draga Claudia, "eforturile" sunt rasplatite cu 4 puncte si bineinteles mii de multumiri, atat tie cat si celorlalte colege !! ;)"
10 hrs

concentratia cunoscuta de SO2

Zero/Span
Part of the QAQC process for many of the continuous analyzers is the daily zero/span cycle. Each day, usually in the early hours of the morning, the normal sample port is closed to the analyzer and, for about twenty minutes, zero air is fed to the analyzer. Zero air is air from a cylinder or air that has been filtered so that it contains almost no pollutant. The analyzer should read zero. If the analyzer does not read zero, then the data is adjusted by a corresponding amount.
This common procedure is known as baseline correction. After the zero air comes the span gas. Span gas is air with a known high concentration of pollutant. The analyzer should output a reading that is within ten percent of the known concentration. If it does not, then the analyzer must be calibrated. After the zero and span gases have been passed through the analyzer, the sample port is again opened. About ten minutes is allowed for the analyzer to recover before readings are accepted again as real.
http://wbea.org/content/view/24/63/#zero

Span-Gas - A gas of known concentration that is used routinely to calibrate the output level of an analyzer. See “Calibration Check Standard”.
http://www.env.gov.bc.ca/wsd/data_searches/field_sampling_ma...
Suppression ratio: The ratio of the lower-range value to the span. If range is 20-100 and therefore span is 80 and LRV is 20, the suppression ratio is 20/80 = 0.25 or 25%.
http://www.omega.com/literature/transactions/volume3/glossar...
In realitate, insa, pentru etalonarea analizoarelor se folosesc butelii etalon cu o concentratie cunoscuta de gaz. Pentru gazele de ardere, amestecurile de gaze etalon cele mai bune au o eroare relativa a concentratiei etalon de gaz de 1%.
http://www.emisii.ro/Erori monitorizare emisii, recomandari....
Something went wrong...

Reference comments

23 hrs
Reference:

SO2 span

Degeaba, nu sunt împăcată cu traduecrea lui "span" în "span"! Îmi cer scuze Claudiei, nu vreau să o supăr însă nu e corect să lăsăm aşa lucrurile.
Argumente:
1) Vezi http://www.proz.com/kudoz/english_to_romanian/mechanics_mech...
2) Range of interval
Range
The range of the interval [a; b] is the difference b - a and is denoted by r[a; b].
NOTE The term “span” is sometimes used for this concept. (pag. 12)

3) Masurare SO2
Domeniu de m`sur`: 0..2000 ppm
Rezolutie: 1 ppm
Precizie: ±10 ppm (<150 ppm)
±5 % dvm (Δ150 ppm)
http://www.afriso.ro/multilyzer.html

Personal, aş opta pt. "interval/domeniu de măsurare SO2".


--------------------------------------------------
Note added at 23 ore (2009-12-14 17:30:10 GMT)
--------------------------------------------------

http://www.afriso.ro/multilyzer.html

--------------------------------------------------
Note added at 23 ore (2009-12-14 17:30:43 GMT)
--------------------------------------------------

http://www.afriso.ro/multilyzer.html

--------------------------------------------------
Note added at 23 ore (2009-12-14 17:31:16 GMT)
--------------------------------------------------

Nu pot posta linkurile:(

--------------------------------------------------
Note added at 23 ore (2009-12-14 17:35:02 GMT)
--------------------------------------------------

http://www.bipm.org/utils/common/documents/jcgm/JCGM_200_200...

si

http://www.afriso.ro/multilyzer.html

--------------------------------------------------
Note added at 23 ore (2009-12-14 18:14:53 GMT)
--------------------------------------------------

Sper ca acum să pot posta:
http://www.bipm.org/utils/common/documents/jcgm/JCGM_200_200...
şi
http://www.afriso.ro/multilyzer.html
Note from asker:
Draga Cristina, nu cred ca se supara nimeni, orice parere este pretioasa si ajuta mult la stabilirea celei mai bune variante finale ;)
Something went wrong...